Motor Control

Serial Command Reference

Commands are issued via serial terminal. All inputs are plain ASCII and case-insensitive.
Parameters are space-separated. All commands must be followed by \r\n.
Valid commands are acknowledged with OK. Invalid commands return !.
All connections are to be made at 9600 baud through the onboard mini-USB port.


Settings Control

Command Description Valid Range Example
SETRPM <rpm> Set the PID target RPM 60–300 SETRPM 150
SETKP <value> Set the PID Kp value 0.0–100.0 SETKP 0.5
SETKI <value> Set the PID Ki value 0.0–100.0 SETKI 0.07
SETKD <value> Set the PID Kd value 0.0–100.0 SETKD 0.0
SETCURRENTLIM <mA> Set motor current limit in mA 100–1000 SETCURRENTLIM 300
SETCUTOFF <0|1> Enable (1) or disable (0) current cutoff protection 0 = off, 1 = on SETCUTOFF 1
SETRESTART <0|1> Enable (1) or disable (0) auto-restart on stall 0 = off, 1 = on SETRESTART 1

Diagnostics and Output

Command Description
SHOW Print current configuration and status
DUMPLOG Dump all logged data in CSV format

Maintenance

Command Description
RESETCONFIG Reset all settings to factory defaults and clear log

Firmware Details

FRAM Memory Map (0x0000 – 0x03EF Reserved)

The first ~1 KB of FRAM is reserved for persistent configuration and metadata.

Offset Size Name Type Description
0x0000 MotorControllerSettings Packed structure (see fields below)

Structure layout:

Field Size Type Description
power_cycle_count 2 uint16_t Number of times the device has booted
log_head_index 2 uint16_t Index of the next log entry
current_limit_ma 2 uint16_t Maximum motor current before shutdown (mA)
pid_kp 4 float Proportional PID gain
pid_ki 4 float Integral PID gain
pid_kd 4 float Derivative PID gain
setpoint_rpm 2 uint16_t Target motor RPM
restart_enabled 1 uint8_t Auto-restart on stall (0 = off, 1 = on)
current_cutoff_enabled 1 uint8_t Current cutoff protection (0 = off, 1 = on)

Log Entry Format (starting at 0x03F0)

Each log entry is 16 bytes and stored in a circular buffer.

Field Size Type Description
timestamp_s 4 uint32_t Timestamp (seconds since boot)
rpm 2 int16_t Current measured RPM
current_ma 2 uint16_t Motor current (mA)
temp_x10 2 int16_t Controller temperature in °C ×10
battery_mv 2 uint16_t Battery voltage (mV)
power_cycles 2 uint16_t Power cycle count at time of log
flags 1 uint8_t Reserved for future use
reserved[2] 2 Padding / future expansion

Logging Behavior


Note: This documentation matches firmware version 1.0 as of the latest revision. See SHOW output for runtime verification.  

Firmware Checkout Procedure

This checklist can be used to verify proper operation and no regressions after the device firmware has been modified.

Tests are designed to verify that each serial command behaves correctly, including:

Test Outcome Legend

SETRPM (Valid range: 60–300)

Test Case Expected Outcome
SETRPM (failure expected) – too few arguments
SETRPM 100 200 (failure expected) – too many arguments
SETRPM 20 (failure expected) – below valid range
SETRPM 400 (failure expected) – above valid range
SETRPM 150 (success expected) – valid RPM set
Power cycle, then SHOW (success expected) – verify setpoint_rpm = 150
SETRPM qwertyuiopasdfghjklzxcvbnm1234567890 (failure expected) – invalid input format

SETKP, SETKI, SETKD (Valid range: 0–100)

Each command should be tested independently. Example below is for SETKP.

Test Case Expected Outcome
SETKP (failure expected) – too few arguments
SETKP 1 2 (failure expected) – too many arguments
SETKP -1 (failure expected) – below valid range
SETKP 101 (failure expected) – above valid range
SETKP 0.05 (success expected) – gain accepted
Power cycle, then SHOW (success expected) – verify value persisted
SETKP ???!!!@@@### (failure expected) – invalid input format

Repeat for:

SETCURRENTLIM (Valid range: 100–1000)

Test Case Expected Outcome
SETCURRENTLIM (failure expected) – too few arguments
SETCURRENTLIM 1 2 3 (failure expected) – too many arguments
SETCURRENTLIM 50 (failure expected) – below range
SETCURRENTLIM 1500 (failure expected) – above range
SETCURRENTLIM 500 (success expected) – valid current limit
Power cycle, then SHOW (success expected) – verify value persisted
SETCURRENTLIM spamspamspamspam (failure expected) – non-numeric input

SETCUTOFF <0|1>

Test Case Expected Outcome
SETCUTOFF (failure expected) – too few arguments
SETCUTOFF 1 2 (failure expected) – too many arguments
SETCUTOFF 2 (failure expected) – invalid value
SETCUTOFF -1 (failure expected) – invalid value
SETCUTOFF 0 (success expected) – disable cutoff
Power cycle, then SHOW (success expected) – verify persisted
SETCUTOFF 1 (success expected) – enable cutoff
Power cycle, then SHOW (success expected) – verify persisted
SETCUTOFF lolnope (failure expected) – non-numeric input

SETRESTART <0|1>

Same validation process as SETCUTOFF.

SHOW

Test Case Expected Outcome
SHOW (success expected) – prints current settings in readable format

DUMPLOG

Test Case Expected Outcome
DUMPLOG (success expected) – prints data log (empty or not)

RESETCONFIG

Test Case Expected Outcome
RESETCONFIG (success expected) – settings reset to defaults, log cleared
Power cycle, then SHOW (success expected) – default settings confirmed
Verify power_cycle_count == 1 (success expected) – after reset

HELP

Test Case Expected Outcome
HELP (success expected) – lists available commands and usage

Junk Input Stress Test

Test Case Expected Outcome
AJSDHASKJDH1238127312##@@!!@!!()()()()_+_)(*&^%$#@!~ (observe) – no crash; next command works
LONGSTRINGOFNONCOMMANDSANDCHARSANDNUMBERS1234567890... (observe) – no buffer overflow
Send partially typed command (observe) – system should not crash or become unresponsive

Power Cycle Verification

After each setting command that returns OK:

Step Expected Outcome
Power off and wait 10 seconds -
Power on and run SHOW (success expected) – changed setting persisted
Confirm power_cycle_count incremented (success expected)

Over Current Protection and Restart Functionality

| Step | Expected Outcome | |——|——————| | With the current limit set to 100 mA stall the motor and verify shutdown executes | (success expected) | | With the restart function enabled the motor should restart in approximately 1 minute | (success expected) | | With the restart function disabled the system should stay shutdown - wait for 5 minutes to verify | (success expected) |

RPM Verification

| Step | Expected Outcome | |——|——————| | Set the RPM to 210 and verify speed is reached within 60 seconds and maintained | (success expected) | | Decrease the voltage driving the system by 2 VDC and verify the RPM recovers within 60 seconds | (success expected) | | Reset the driving voltage and verify the RPM recovers within 60 seconds | (success expected) |

Cleanup

After completing the firmware testing, ALWAYS reset the system to the defaults for operation. Adjust any custom settings as desired.